A Novel Non-invasive Method for the Detection of Elevated Intra-compartmental Pressures of the Leg
Published on: May 31, 2019
Noninvasive Technique for Measuring Central Venous and Arterial Pressure Using Controlled Compression Sonography
Hiroshi Tomoeda1, Kentaro Sawada2, Shingo Chihara3
1Department of Cardiovascular Surgery, Chikugo City Hospital, Chikugo, Fukuoka, Japan.
A novel noninvasive controlled compression sonography system accurately measures venous pressures beyond conventional limits. While effective for venous pressure, further research is needed to confirm its clinical utility for arterial pressure monitoring.
Area of Science:
- Medical Devices
- Diagnostic Technology
- Cardiovascular Monitoring
Background:
- Existing noninvasive devices for measuring venous pressure have limited resolution (up to 50 mmHg).
- There is a need for advanced noninvasive methods to measure higher venous and arterial pressures accurately.
Purpose of the Study:
- To develop and validate an innovative noninvasive controlled compression sonography system.
- To overcome the resolution limitations of current devices for pressure measurement.
- To assess the system's capability for measuring central venous, peripheral venous, and brachial artery pressures.
Main Methods:
- Utilized a noninvasive controlled compression sonography device capable of measuring pressures exceeding 200 mmHg.
- Conducted an in vitro study for device accuracy validation.
- Compared sonography measurements with conventional methods for venous and arterial pressures.
- Performed regression analyses to determine correlations between different measurement techniques.
Main Results:
- The developed system measured venous and arterial pressures in 3 to 15 seconds.
- Significant statistical correlations were found between the sonography system and conventional measurements.
- The correlation for arterial pressure measurements was relatively low, indicating a need for further refinement.
Conclusions:
- The novel sonography system successfully measured venous pressures noninvasively, exceeding the capabilities of conventional methods.
- The system shows potential for measuring systolic arterial pressure.
- Further clinical studies are required to establish the efficacy of this device for arterial pressure monitoring.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Venous Thrombosis II: Clinical Manifestations and Diagnostic Studies
Assessment of blood pressure in brachial artery(two-step method)