Reference dosimetry of modulated and dynamic photon beams
Hugo Bouchard1,2,3
1Département de physique, Université de Montréal, Complexe des sciences, 1375 Avenue Thérèse-Lavoie-Roux, Montréal, Québec H2V 0B3, Canada.
Physics in Medicine and Biology
|January 13, 2021
Summary
Intensity-modulated radiotherapy (IMRT) lacks calibration guidelines. This paper reviews nonstandard photon beam dosimetry and discusses implementing new IAEA-AAPM formalism for dynamic IMRT beams.
Area of Science:
- Medical Physics
- Radiation Oncology
- Radiotherapy Dosimetry
Background:
- Intensity-modulated radiotherapy (IMRT) has advanced photon beam delivery since the late 1980s.
- Techniques like helical tomotherapy and volumetric modulated arc therapy are standard in modern radiotherapy.
- Despite widespread clinical use, established guidelines for calibrating dynamic modulated photon beams are still lacking.
Purpose of the Study:
- To conduct a literature review on the dosimetry of nonstandard photon beams, including dynamic deliveries.
- To discuss the potential benefits and challenges of implementing the IAEA-AAPM formalism for dynamic photon beams.
- To highlight the need for updated dosimetry protocols for advanced radiotherapy techniques.
Main Methods:
- Literature review focusing on nonstandard photon beam dosimetry and calibration challenges.
- Analysis of the IAEA-AAPM formalism proposed by Alfonso et al. (2008).
- Discussion of the implications of IAEA Code of Practice TRS-483 and its limitations for dynamic beams.
Main Results:
- A significant gap exists in established guidelines for the calibration of dynamic modulated photon beams.
- The IAEA-AAPM formalism offers a new approach to reference dosimetry for nonstandard beams.
- IAEA TRS-483 addresses small static beams, but the broader formalism for dynamic beams requires clinical implementation.
Conclusions:
- Implementing the IAEA-AAPM formalism for dynamic photon beams represents a paradigm shift in radiotherapy calibration.
- Addressing the challenges of this implementation is crucial for accurate and safe IMRT delivery.
- Further research and clinical adoption are needed to establish robust dosimetry protocols for advanced radiotherapy techniques.