Experimental determination of thermal neutron fluence around Elekta Versa HD linear accelerator for various photon
R Vysakh1,2, M M Musthafa3, C V Midhun3
1Department of Physics, Noorul Islam Centre for Higher Education, Kumaracoil- 629 180, Kanyakumari District, Tamil Nadu, India.
Abstract:
A complex neutron spectrum generated along with a useful photon beam imposes an additional radiation protection risk around medical linear accelerators (linac). The thermal neutron component of this complex neutron spectrum formed during different photon modes of operation of Elekta Versa HD linac has been quantified using Indium foil activation technique. The thermal neutron fluence (Φ th ) at isocenter for 15 MV, 10 MV and 10 MV FFF beams was found to be 2.45 × 105, 4.35 × 104 and 3.2 × 104 neutrons cm-2 Gy-1, respectively. The analysis shows a reduction in the Φ th as the flattening filter is being taken out from the beam path. A negative correlation in Φ th with respect to field size has been observed with an average 18% reduction in Φ th per monitor units as field size changes from 10 cm × 10 cm to 40 cm × 40 cm. For particular field size and photon energy, Φ th was found to be uniform across the patient plane. From the measured gamma ray spectrum inside the treatment room six major isotopes have been identified which were 122Sb, 187W, 82Br, 56Mn, 24Na and 28Al.
More Related Videos
08:31Sample Preparation and Experimental Design for In Situ Multi-Beam Transmission Electron Microscopy Irradiation Experiments
Published on: June 27, 2022
10:24Neutron Radiography and Computed Tomography of Biological Systems at the Oak Ridge National Laboratory's High Flux Isotope Reactor
Published on: May 7, 2021