Related Experiment Video
Updated: Nov 21, 2025

Author Spotlight: Improving Percutaneous Vertebroplasty Surgical Accuracy and Efficiency Through Advanced Puncture Techniques
Published on: August 9, 2024
Robot-assisted percutaneous vertebroplasty under local anaesthesia for osteoporotic vertebral compression fractures:
Bin Shi1, Lei Hu2, Hailong du1
1Department of Orthopedics, Chinese PLA General Hospital, Beijing, China.
Research And Design:
A retrospective, clinical, non-randomized, controlled study.
Objective:
To evaluate the application of a new spinal navigation robot on percutaneous vertebroplasty (PVP) under local anaesthesia.
Methods:
The result of inserting the puncture needle into the simulated pedicle was observed in vitro. Thirty patients with thoracolumbar fractures were enrolled. The basic data, operation-related data and clinical effect scores were recorded. The learning curve of robot-guided surgery was analysed.
Results:
The maximum yaw angle of the puncture needle implanting into the simulated pedicle is 0.5°. The operation time (33.93 ± 5.97 min vs. 53.75 ± 14.08 min, p = 0.000) and the average X-ray exposure time (31.43 ± 4.93 s vs. 54.69 ± 2.15 s, p = 0.000) was significantly less in the robot group. The surgeons quickly mastered the technique of robot-guided pedicle puncture after three surgeries.
Conclusions:
Robot-assisted PVP could performed under local anaesthesia with the new robot.