Protocol for Relative Hydrodynamic Assessment of Tri-leaflet Polymer Valves
Published on: October 17, 2013
Proper Orthogonal Decomposition Analysis of the Flow Downstream of a Dysfunctional Bileaflet Mechanical Aortic Valve
Ahmed Darwish1,2, Giuseppe Di Labbio3,4, Wael Saleh3,5
1Laboratory of Cardiovascular Fluid Dynamics, Department of Mechanical, Industrial and Aerospace Engineering, Concordia University, 1455 de Maisonneuve Blvd. W, Montréal, QC, H3G 1M8, Canada. lcfd@encs.concordia.ca.
Purpose:
Aortic valve replacement remains the only viable solution for symptomatic patients with severe aortic valve stenosis. Despite their improved design and long history of successful operation, bileaflet mechanical heart valves are still associated with post-operative complications leading to valve dysfunction. Thus, the flow dynamics can be highly disturbed downstream of the dysfunctional valve.
Methods:
In this in vitro study, the flow dynamics downstream of healthy and dysfunctional bileaflet mechanical heart valves have been investigated using particle image velocimetry measurements. Proper orthogonal decomposition of the velocity field has been performed in order to explore the coherent flow features in the ascending aorta in the presence of a dysfunctional bileaflet mechanical heart valve.
Results:
The ability of proper orthogonal decomposition derived metrics to differentiate between heathy and dysfunctional cases is reported. Moreover, reduced-order modeling using proper orthogonal decomposition is thoroughly investigated not only for the velocity field but also for higher order flow characteristics such as time average wall shear stress, oscillatory shear index and viscous energy dissipation.
Conclusion:
Considering these results, proper orthogonal decomposition can provide a rapid binary classifier to evaluate if the bileaflet mechanical valve deviates from its normal operating conditions. Moreover, the study shows that the size of the reduced-order model depends on which flow parameter is required to be reconstructed.
More Related Videos
11:16In vitro Assessment of Aortic Regurgitation Using Four-Dimensional Flow Magnetic Resonance Imaging
Published on: February 25, 2022
11:00Experimental Investigation of Secondary Flow Structures Downstream of a Model Type IV Stent Failure in a 180° Curved Artery Test Section
Published on: July 19, 2016
Related Concept Videos
Aortic Regurgitation I: Introduction
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Mitral Valve Prolapse I: Introduction
Aortic Regurgitation II: Clinical Features and Diagnostic Tests
Thin-Walled Hollow Shafts