Equipment Capability Measurement of Laboratory Freeze-Dryers: a Comparison of Two Methods
Jayasree M Srinivasan1, Gregory A Sacha2, Vaibhav Kshirsagar3
1Pharmaceutical R&D, Baxter Healthcare, 927 S. Curry Pike, Bloomington, Indiana, 47403, USA. jayasree_srinivasan@baxter.com.
Abstract:
The objective of this investigation was to evaluate two methods for measuring the maximum sublimation rate that a freeze-dryer will support-the minimum controllable pressure method and the choke point method. Both methods gave equivalent results, but the minimum controllable pressure method is preferred, since it is easier, faster, and less subjective. The ratio of chamber pressure to condenser pressure corresponding to the onset of choked flow was considerably higher in this investigation (up to about 20:1) than in previously published reports. This ratio was not affected by the location of the pressure gauge on the condenser; that is, on the foreline of the vacuum pump versus on the body of the condenser itself. The total water loss due to sublimation as measured by tunable diode laser absorption spectroscopy was consistently within 5% of gravimetrically determined weight loss, regardless of whether the measurement took place during choked versus non-choked process conditions.
More Related Videos
12:52Tandem High-pressure Freezing and Quick Freeze Substitution of Plant Tissues for Transmission Electron Microscopy
Published on: October 13, 2014
07:57Quantitative Detection of Trace Explosive Vapors by Programmed Temperature Desorption Gas Chromatography-Electron Capture Detector
Published on: July 25, 2014