Related Experiment Video
Updated: Nov 20, 2025

Guide Wire Assisted Catheterization and Colored Dye Injection for Vascular Mapping of Monochorionic Twin Placentas
Published on: September 5, 2011
Entrapment of diagnostic catheter within Advisor HD grid mapping catheter
Philip L Mar1, Luke Chong1, Arturo Perez1
1Division of Cardiology, Saint Louis University, Saint Louis, Missouri, USA.
Abstract:
The potential for catheter entanglement with the HD Grid mapping catheter is explicitly stated in the manufacturer's product manual. A case of an entrapped 6 French quadripolar diagnostic catheter within an HD Grid mapping catheter is presented. We discuss the diagnosis, management, and resolution of this complication in our patient. The patient's arrhythmia was successfully eliminated, and no vascular complication in the postprocedural setting nor arrhythmia recurrence at follow-up were observed. Strategies to prevent and safely manage this complication, while salvaging access, are also discussed.
More Related Videos
Related Concept Videos
Cardiac Catheterization III: Left Heart Catheterization
Cardiac Catheterization I: Pre-Procedure Overview