Related Experiment Video
Updated: Jul 13, 2026

Behavioral Assessment of Hearing in 2 to 4 Year-old Children: A Two-interval, Observer-based Procedure Using Conditioned Play-based Responses
Published on: January 23, 2017
Reference thresholds for the ER-3A insert earphone
L A Wilber1, B Kruger, M C Killion
1Program in Audiology, Northwestern University, Evanston, Illinois 60208.
Abstract:
Several recent studies have demonstrated that the ER-3A insert earphone may sometimes be directly substituted, without recalibrating, for a TDH-39/MX-41AR earphone. However, most available data have not been reduced to a form suitable for establishing a revised estimate of the reference threshold levels. This article reports such a data analysis performed on the results of five recent studies. The mean data from the five studies are typically within 1 dB of the provisional reference threshold SPLs given by the ER-3A manufacturer for calibration in a (HA-1) 2-cc coupler. After converting the mean data to equivalent Zwislocki-coupler-type ear simulator SPLs at each of the reported audiometric frequencies (125, 250, 500, 1000, 2000, 3000, 4000, 6000, and 8000 Hz), agreement within 1.5 dB was seen with the revised estimate of minimum audible pressures given by Killion [J. Acoust. Soc. Am. 63, 1501-1508 (1978)]. Either the manufacturer's provisional SPLs or the average results from this study may be used with little noticeable difference for most purposes.