Related Experiment Video
Updated: Nov 19, 2025

Fabrication and Testing of Photonic Thermometers
Published on: October 24, 2018
Calibration and Testing of Small High-Resolution Transition Edge Sensor Microcalorimeters with Optical Photons
Felix T Jaeckel1, C V Ambarish1, Haoran Dai1
1Department of Physics, University of Wisconsin-Madison, Madison, WI 53706 USA.
Abstract:
Pulses of narrow line-width optical photons can be used to calibrate and test sub-2 eV full-width at halfmaximum (FWHM) energy resolution transition-edge sensor (TES) microcalorimeters at low energies (< 1 keV), where it is very challenging to obtain X-ray calibration lines comparable to (or narrower than) the detector resolution. This scheme depends on the ability to resolve the number of 3 eV photons in each pulse, which we have recently demonstrated up to photon numbers of about 300. At LTD-18 we showed preliminary results obtained with this technique on a 0.25 eV baseline resolution TES microcalorimeter designed for the ultra-high-resolution subarray of the Lynx mission. The line-shape was well described by a simple Gaussian. However, the difficulty of delivering photons to the small 46 microm square absorbers resulted in a large thermal crosstalk signal, whose random nature is expected to rapidly degrade the observed energy resolution towards higher photon numbers/energies. We have since improved the coupling between the optical fiber and the TES absorber and report here our current results.
Related Concept Videos
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Calorimetry

