Related Experiment Video
Updated: Nov 18, 2025

Bergmeyer Glucose Quantification for Microbiological Samples
Published on: January 17, 2025
Evaluation of Trueness and Precision of a Bench-Top Laboratory Glucose Analyzer Using Reference Materials
Annette Baumstark1, Stefan Pleus1, Matthias Hartwig2
1Institut für Diabetes-Technologie, Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm, Ulm, Germany.
Abstract:
SUPER GL compact is a bench-top analyzer for glucose, lactate, and hemoglobin concentrations. Glucose measurements in the biosensor are based on an enzymatic-amperometric reaction of glucose with glucose oxidase.In this study, trueness and precision were assessed with Standard Reference Material 965b (National Institute of Standards and Technology, Gaithersburg, MD) for 2 SUPER GL compact (S1 and S2) and 1 YSI 2300 STAT Plus (Y) device, using a protocol based on CLSI EP05-A3.Precision was similar among S1, S2, and Y. S1 and S2 exhibited negative bias at low concentrations and positive bias at high concentrations, whereas Y showed negative bias that increased with higher concentrations. Overall, SUPER GL compact's performance was comparable to that of YSI 2300 STAT Plus.