Comparative Performance Analysis of Manual and Updated Optional Calibration Algorithms for the CareSens Air CGM
Nina Jendrike1, Manuel Eichenlaub1, Manuela Link1
1Institut für Diabetes-Technologie, Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm, Ulm, Germany.
The updated CareSens Air continuous glucose monitoring (CGM) system shows improved accuracy and reliability compared to its previous version. This enhanced CGM system is suitable for nonadjunctive use in diabetes management.
Area of Science:
- Biomedical Engineering
- Medical Devices
- Diabetes Technology
Background:
- The CareSens Air continuous glucose monitoring (CGM) system (CSAir) initially required manual calibration and had a 2-hour warm-up period.
- Product development led to an updated algorithm reducing the warm-up time to 30 minutes and making calibrations optional.
Purpose of the Study:
- To compare the performance of the CareSens Air CGM system using its original manual calibration algorithm versus the updated algorithm.
- To evaluate the accuracy, reliability, and user satisfaction of the updated CSAir system.
Main Methods:
- Thirty adults with diabetes participated, wearing three CSAir sensors for 15 days each.
- In-clinic sessions involved frequent capillary comparator measurements and glucose level manipulation.
- Accuracy was assessed using mean absolute relative difference (MARD) and 20/20 agreement rates (AR).
Main Results:
- The updated algorithm demonstrated superior accuracy with a 93.9% 20/20 AR and 8.7% MARD, compared to 90.1% and 9.9% for the manual algorithm.
- High clinical accuracy was confirmed by Diabetes Technology Society Error Grid analysis (92.4% in zone A for updated algorithm).
- The system showed good sensor wear time, positive user satisfaction, and no safety concerns.
Conclusions:
- Both CSAir algorithms performed reliably, with the updated version showing enhanced accuracy.
- The CareSens Air CGM system, particularly with the updated algorithm, is suitable for nonadjunctive diabetes management.
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Errors occurring during blood pressure monitoring
Several factors...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...