Increased incidence of electrical abnormalities in a pacemaker lead family
Evan Adelstein1, Lane Zhang1, Haider Nazeer1
1Department of Medicine, Division of Cardiology, Albany Medical Center, Albany, New York, USA.
Introduction:
Several recent studies have raised concern about noise detections on Tendril pacemaker leads, which may represent insulation failure or algorithm-driven overreporting of physiologic signals.
Methods:
We identified all pacemaker leads actively followed at Albany Medical Center, of which 1111 leads (262 Abbott Tendril, 576 Medtronic CapSure, 30 Fineline, 195 Ingevity, 48 Dextrus) in 703 patients were included in this observational study. Electrical abnormalities, comprising low-impedance measurements <200 Ω and repeated noise detections, were catalogued, as was initial management and subsequent need for surgical intervention.
Results:
During 54 months median follow-up (interquartile range 24-105), 63 leads (5.7%) demonstrated electrical abnormalities, including low impedances in 21 and noise in 59. Tendril leads manifested abnormalities most frequently (n = 50; 19.1%) compared with CapSure (n = 9; 1.6%), Fineline (n = 0), Ingevity (n = 0), and Dextrus (n = 4; 8.3%) leads. The risk of abnormalities was significantly higher in Tendril leads (HR 9.6, 95% CI 5.2-17.6; p < .001). Low impedances were measured on 19 Tendril leads, a significantly higher risk than on other leads (HR 23.8, 95% CI 5.5-102.1; p < .001). Although observation and reprogramming sensitivity were the initial management strategy for 45 and 7 leads, respectively, 18 ultimately required surgical intervention, including 15 Tendrils. No electrical abnormalities were observed in 12 non-Tendril leads attached to Abbott devices compared with 48 of 252 Tendrils attached to Abbott devices (log-rank p = .035).
Conclusion:
Tendril leads demonstrate significantly higher risk of repeated low impedances and noise compared to other manufacturers' models, raising concern that these findings reflect early insulation failure. Increased scrutiny is warranted.
Related Concept Videos
Mechanism of Cardiac Arrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Electrophysiology of Normal Cardiac Rhythm
Dysrhythmias III: Characteristics of Dysrhythmias