Portable gas chromatograph calibration with gases of varying viscosities
Vera Tschedanoff1, Tilman J Schildhauer1, Serge M A Biollaz1
1General Energy Research Department, Paul Scherrer Institut, Villigen PSI, CH-5232, Switzerland.
Abstract:
The quantitative analysis of gas mixtures with gas chromatography is based on the calibration with certified standards and the determination of a response relationship for each species by regression analysis. The conventional assumption of a constant amount-of-substance injected onto the column, the sample size, for all standards analyzed represents one of the largest sources of uncertainies in this analysis technique. For systems using time-based microinjectors, the sample size injected onto the column is determined by the opening time of a pneumatically actuated microvalve and therefore depends upon the gas velocity developed in the injector's microchannel. For this reason the sample size is not necessarily constant for a given opening time, but strongly correlates - according to fluid mechanic theory - with the gas mixture's dynamic viscosity. Neglecting these sample size variations leads to errors in the analysis up to 30%, depending on the diversity of the standards' viscosities, especially in processes with strongly changing hydrogen contents. A mathematical correction exploiting the inverse proportionality of the sample size and the sample's dynamic viscosity normalizes the injection amounts and minimizes the influence of the sample variations on the calibration regression. The data analysis based on the corrected normalized calibration is no longer viscosity dependent and can be applied to gas mixtures of any composition.
More Related Videos
08:22Nitrogen Compound Characterization in Fuels by Multidimensional Gas Chromatography
Published on: May 15, 2020
08:37Measurement of H2S in Crude Oil and Crude Oil Headspace Using Multidimensional Gas Chromatography, Deans Switching and Sulfur-selective Detection
Published on: December 10, 2015
Related Concept Videos
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...