Radiation dose, contrast enhancement, image noise and heart rate variability of ECG-gated CT volumetry using 3D
1Department of Radiology and Research Institute of Radiology, University of Ulsan College of Medicine, Asan Medical Center, 88, Olympic-ro 43-gil, Songpa-gu, Seoul 05505, Republic of Korea.
Objective:
To compare radiation dose, contrast enhancement, image noise and heart rate variability in electrocardiography (ECG)-gated computed tomography (CT) ventricular volumetry using a three-dimensional (3D) threshold-based segmentation between the conventional single scan and dual focused scan methods in patients with congenital heart disease.
Methods:
After matching for age, sex, heart rate during the CT examination, and tube voltage, 96 patients (age range, 7 - 36 years; male:female = 63:33) who underwent ECG-gated CT volumetry using a 3D threshold-based segmentation, were divided into 32 patients who underwent a conventional single scan (group 1) and 64 who underwent dual focused scans (group 2). CT radiation dose, contrast enhancement, image noise, and heart rate variability were compared between the two groups.
Results:
Volume CT dose index, dose-length product, and effective dose estimates, in group 1 were significantly higher than those in group 2 (28.4 ± 24.6 mGy vs. 9.7 ± 4.5 mGy, 636.5 ± 572.9 mGy cm vs. 379.5 ± 192.4 mGy cm, 8.9 ± 8.0 mSv vs. 5.3 ± 2.7 mSv, 8.2 ± 6.4 mSv vs. 5.0 ± 2.2 mSv, respectively; p values < 0.019). No significant differences in right and left ventricular contrast enhancement were found between the two groups (470.2 ± 152.4 HU vs. 481.0 ± 139.2 HU and 428.6 ± 97.8 HU vs. 429.3 ± 106.7 HU, respectively; p values > 0.05). Image noise measured in the air showed no significant differences between groups 1 and 2 (5.6 ± 1.9 HU vs. 5.5 ± 1.1 HU; p > 0.05). The proportion of heart rate variability ≥ 20 beats per minute in group 1 (15.6 %, 5/32) was significantly higher than that in group 2 (3.1 %, 2/64; p < 0.040).
Conclusion:
Compared with the conventional single scan method, the dual focused scan method can provide a lower radiation dose with comparable contrast enhancement and image noise for ECG-gated CT ventricular volumetry using a 3D threshold-based segmentation in patients with congenital heart disease.
More Related Videos
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
11:38Voluntary Breath-hold Technique for Reducing Heart Dose in Left Breast Radiotherapy
Published on: July 3, 2014
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography