Related Experiment Video
Updated: Nov 16, 2025

Author Spotlight: Segmentation and VR for Advanced Neurovascular Interventions
Published on: April 5, 2024
Comparison of three-dimensional (3D) endovision system versus ultra-high-definition 4K endovision system in minimally
Rajinder Parshad1, Shashikiran Nanjakla Jayappa2, Hemanga Kumar Bhattacharjee1
1Department of Surgical disciplines, All India Institute of Medical Sciences, Ansari Nagar, New Delhi, 110029, India.
Background:
Experimental work comparing 3-Dimensional (3D) and 4K ultra-high-definition endovision system (4K) indicates that the latter with double the resolution of standard 2D high-definition systems may provide additional visual cues to compensate for the lack of stereoscopic vision. There is paucity of studies comparing 3D and 4K system in clinical settings. This randomized study compares 3D and 4K systems in three laparoscopic procedures of increasing complexity.
Methods:
139 patients undergoing laparoscopic cholecystectomy (60 patients), transabdominal preperitoneal (TAPP) repair (49 patients) and laparoscopic Heller's cardiomyotomy with anti- reflux procedure (30 patients) between May 2018 and February 2020 were randomized to undergo surgery using either 3D or 4K systems. Primary objective was to measure total operative time. Secondary objectives were to compare workload perceived by surgeons using SURG-TLX and surgeon satisfaction score. Timings of key surgical steps and peri-operative course of the patients was also recorded. Data were analyzed using Stata Corp. 2015.
Results:
Patients undergoing surgery with 3D and 4K systems were comparable in their clinical and demographic profiles. The mean total operative time in 3D and 4K groups was comparable in cholecystectomy (52.7 vs 56.2, p = 0.50), TAPP (63.8 vs 69.6, p = 0.25) and Heller's cardiomyotomy (124.7 vs 143.3, p = 0.14) with faster hiatal dissection in 3D group (8 min, p = 0.02). Operative time was better in patients undergoing Heller's myotomy with Angle of His accentuation with 3D by 28 min (p = 0.03). Total workload was similar in 3D and 4K groups in all the procedures but mental & physical demand was lower in 3D group in Heller's cardiomyotomy (p = 0.03, p = 0.01), Surgeon satisfaction score was comparable in all three procedures.
Conclusion:
Overall, 3D HD and 4K systems are comparable in performing laparoscopic cholecystectomy, TAPP and Heller's Cardiomyotomy. Hiatal dissection time, mental and physical task load was better with 3D in Heller's Cardiomyotomy.
More Related Videos
05:12Robotized Testing of Camera Positions to Determine Ideal Configuration for Stereo 3D Visualization of Open-Heart Surgery
Published on: August 12, 2021
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
Related Concept Videos
Endoscopic Procedures III: Video Capsule Endoscopy
Endoscopic Studies I: Bronchoscopy and Thoracoscopy
Bronchoscopy
Description
Bronchoscopy is a procedure that involves direct visualization of the larynx, trachea, and bronchi for diagnostic and therapeutic purposes. A flexible fiber optic or rigid bronchoscope is used to carry out the procedure. The fiber-optic bronchoscope is more frequently used due...
Endoscopic Procedures IV: Sigmoidoscopy and Laproscopy
Sigmoidoscopy
Sigmoidoscopy is a diagnostic procedure that uses a flexible sigmoidoscope equipped with a light source and camera to examine the rectum and sigmoid colon. The procedure involves inserting the tube through the anus...
Endoscopic Procedures II: Colonoscopy
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to: