Development of a compact X-band linear accelerator system mounted on an O-arm rotating gantry for radiation therapy
Sanghoon Kim1, Yong-Seok Lee1, Geun-Ju Kim1
1Electro-Medical Device Research Center, Korea Electrotechnology Research Institute, Ansan 15588, South Korea.
Abstract:
A compact X-band linear accelerator (LINAC) system equipped with a small and lightweight magnetron was constructed to develop a high-precision image-guided radiotherapy system. The developed LINAC system was installed in an O-ring gantry where cone-beam computed tomography (CBCT) was embedded. When the O-arm gantry is rotated, an x-ray beam is stably generated, which resulted from the stable transmission of radio frequency power into the X-band LINAC system. Quality assurance (QA) tests, including mechanical and dosimetry checks, were carried out to ensure safety and operation performance according to the American Association of Physicists in Medicine's TG-51, 142, an international standard protocol established by accredited institutions. In addition, delivery QA of the radiotherapy planning system was conducted to verify intensity-modulated radiotherapy techniques. Therefore, it was demonstrated that the developed X-band LINAC system mounted on the O-arm gantry proved to be valid and reliable for potential use in CBCT image-guided radiation therapy.