Compatibility and Validation of a Recent Developed Artificial Blood through the Vascular Phantom Using Doppler
Marwan Alshipli1, Mohannad Adel Sayah1, Ammar A Oglat2
1Department of Radiography, Princess Aisha Bint Al-Hussein College of Nursing and Health Sciences, Al-Hussein Bin Talal University, Ma'an, Jordan.
Background:
Doppler technique is a technology that can raise the predictive, diagnostic, and monitoring abilities in blood flow and suitable for researchers. The application depends on Doppler shift (shift frequencies), wherein the movement of red blood cells away from the probe is determined by the decrease or increase in the ultrasound (US) frequency.
Methods:
In this experiment, the clinical US (Hitachi Avious [HI] model) system was used as a primary instrument for data acquisition and test the compatibility, efficacy, and validation of artificial blood (blood-mimicking fluid [BMF]) by color- and motion-mode. This BMF was prepared for use in the Doppler flow phantom.
Results:
The motion of BMF through the vessel-mimicking material (VMM) was parallel and the flow was laminar and in the straight form (regular flow of BMF inside the VMM). Moreover, the scale of color velocity in the normal range at that flow rate was in the normal range.
Conclusion:
The new BMF that is being valid and effective in utilizing for US in vitro research applications. In addition, the clinical US ([HI] model) system can be used as a suitable instrument for data acquisition and test the compatibility, efficacy, and validation at in vitro applications (BMF, flow phantom components).
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...