Blood pressure from the optical Aktiia Bracelet: a 1-month validation study using an extended ISO81060-2 protocol
Anna Vybornova1, Erietta Polychronopoulou2, Arlène Wurzner-Ghajarzadeh2
1Aktiia SA, Neuchâtel ( www.aktiia.com ).
Insights
The Aktiia Bracelet, a novel optical blood pressure monitor, demonstrated high accuracy and precision in a 1-month study. This noninvasive wrist-worn device meets validation criteria for reliable blood pressure measurement in adults.
Area of Science:
- Biomedical Engineering
- Cardiovascular Technology
- Medical Devices
Background:
- Accurate blood pressure (BP) monitoring is crucial for cardiovascular health management.
- Traditional BP measurement methods can be invasive or cumbersome.
- Noninvasive, cuffless BP monitoring offers potential for improved patient convenience and continuous tracking.
Purpose of the Study:
- To evaluate the accuracy and precision of the Aktiia Bracelet, a CE-marked noninvasive optical BP monitor.
- To assess the device's performance over a 1-month period in a clinical setting.
- To determine if the Aktiia Bracelet meets established validation criteria for BP measurement devices.
Main Methods:
- The study involved participants aged 21-65 years, adhering to ISO81060-2:2013 standards adapted for cuffless devices.
- Simultaneous BP measurements were taken using the Aktiia Bracelet and dual-blinded auscultation.
- Optical signals were processed by the Aktiia Bracelet algorithms for quality checks and BP estimation, calibrated using reference measurements.
Main Results:
- Eighty-six participants were analyzed, with mean differences of 0.46 ± 7.75 mmHg for systolic BP (SBP) and 0.39 ± 6.86 mmHg for diastolic BP (DBP).
- The standard deviation of averaged paired differences per subject met ISO81060-2 criteria, with 3.9 mmHg for SBP and 3.6 mmHg for DBP.
- The Aktiia Bracelet demonstrated high accuracy and precision, satisfying key validation criteria.
Conclusions:
- The Aktiia Bracelet achieved overall accuracy meeting ISO81060-2 validation criteria 1 and 2 after initialization and over 1 month.
- The device proved accurate and precise for blood pressure measurement in the sitting position.
- The Aktiia Bracelet is recommended for blood pressure monitoring in the adult population.
Objective:
The objective of this study (NCT04027777) was to assess the accuracy and precision of the Aktiia Bracelet, a CE-marked noninvasive optical blood pressure (BP) monitor worn at the wrist, over a period of 1 month.
Methods:
In this study, participants aged between 21 and 65 years were recruited. The clinical investigation extended the ISO81060-2:2013 standard to the specificities of cuffless devices. Each BP assessment consisted of the simultaneous recording of optical signals with Aktiia Bracelet and double-blinded auscultation by two trained observers in the standard sitting position. The algorithms of Aktiia Bracelet further processed the recorded optical signals to perform a signal quality check and to calculate uncalibrated estimates of systolic BP (SBP) and diastolic BP (DBP). These estimates were transformed into mmHg using a subject-dependent calibration parameter, which was calculated using the first two available reference measurements per subject.
Results:
Eighty-six participants were included in the analysis. The mean and SD of the differences between Aktiia Bracelet estimates and the reference (ISO81060-2 criterion 1) were 0.46 ± 7.75 mmHg for SBP and 0.39 ± 6.86 mmHg for DBP. The SD of the averaged paired difference per subject (ISO81060-2 criterion 2) were 3.9 mmHg for SBP and 3.6 mmHg for DBP.
Conclusion:
After initialization and during 1 month, the overall accuracy of Aktiia Bracelet satisfied validation criteria 1 and 2 of ISO81060-2 in the sitting position. The Aktiia Bracelet can be recommended for BP measurement in the adult population.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
05:51Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
Related Concept Videos
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Pre-Procedural Guidelines for Assessing Blood Pressure
Measurement of Blood Pressure