Perioral Aerosol Sequestration Suction Device Effectively Reduces Biological Cross-Contamination in Dental Procedures
Víctor Lloro1, Maria Laura Giovannoni2, Vicente Lozano-de Luaces3
1Human Anatomy and Embryology Unit, Experimental Pathology and Therapeutics Department, Faculty of Medicine and Health Sciences, Health University of Barcelona campus, L'Hospitalet 08907, Barcelona, Spain.
Objective:
The infection risk during dental procedures is a common concern for dental professionals which has increased due to coronavirus (severe acute respiratory syndrome coronavirus 2) pandemic. The development of devices to specifically mitigate cross-contamination by droplet/splatter is crucial to stop infection transmission. The objective of this study is to assess the effectiveness of a perioral suction device (Oral BioFilter, OBF) to reduce biological contamination spread during dental procedures.
Materials And Methods:
Forty patients were randomized 1:1 to a standard professional dental hygiene treatment with OBF and without. Adenosine triphosphate (ATP) bioluminescence assay was used to evaluate the spread of potential contaminants. The total number of relative light units (RLU) from key dental operatory locations: operator's face-shield, back of the surgical operator's-gloves, patient's safety-goggles, and instrumental table were measured. Percentage contamination reductions between control and OBF were compared.
Statistical Analysis:
Primary outcome, total RLU, was analyzed by comparing the means of logged data, using a two-sided two-sample t-test. Secondary outcomes as RLU of logged data for the different locations were analyzed in the same way. Proportion of patients from whom different locations reported events (clean, acceptable, and failure) were analyzed by using Fisher's exact test.
Results:
For the whole dental environment, RLUs reduction (<150 units) achieved with OBF was 98.4% (97.4-99%). By dental operatory location the reduction in RLUs was from 99.6%, on the operator face-shield, to 83% on instrumental table. The control group reported a very high percentage of failures, (>300) being 100% on the surfaces closer to the patient's mouth and decreasing to 70% on instrumental table. In contrast, the higher failure percentage in the OBF group was found on the patient's goggles (40%), while the operator face-shield showed an absence of contamination.
Conclusion:
OBF device has shown efficient reduction of biological aerosol cross-contamination during dental procedures as proved by ATP-bioluminescence assay. Nevertheless, for maximum safety, its use must be combined with standard protective gear such as goggles, face shield, and surgical gloves.
Related Concept Videos
Suctioning the Nasopharyngeal Airway
Equipment Required
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Transmission-based Precautions II: Airborne and Protective Environment
Airborne precautions:
Use airborne precautions when treating patients known or suspected to have diseases that spread through the air—for example, tuberculosis or measles. These organisms are present in smaller droplets expelled by an infected person and...
Tracheostomy Suctioning II: Procedure
Cardiopulmonary Resuscitation II: ACLS Airway Management
Tracheostomy Suctioning I: Pre-Procedural Steps
Equipment Required
First, gather all necessary equipment: a sterile suction catheter, a sterile disposable container, sterile gloves, a towel or...