Dispersion calibration for the National Ignition Facility electron-positron-proton spectrometers for intense laser
Jens von der Linden1, José Ramos-Méndez2, Bruce Faddegon2
1Lawrence Livermore National Laboratory, Livermore, California 94550, USA.
Abstract:
Electron-positron pairs, produced in intense laser-solid interactions, are diagnosed using magnetic spectrometers with image plates, such as the National Ignition Facility Electron-Positron-Proton Spectrometers (EPPSs). Although modeling can help infer the quantitative value, the accuracy of the models needs to be verified to ensure measurement quality. The dispersion of low-energy electrons and positrons may be affected by fringe magnetic fields near the entrance of the EPPS. We have calibrated the EPPS with six electron beams from a Siemens Oncor linear accelerator (linac) ranging in energy from 2.7 MeV to 15.2 MeV as they enter the spectrometer. A Geant4 traveling-wave optical parametric amplifier of superfluorescence Monte Carlo simulation was set up to match depth dose curves and lateral profiles measured in water at 100 cm source-surface distance. An accurate relationship was established between the bending magnet current setting and the energy of the electron beam at the exit window. The simulations and measurements were used to determine the energy distributions of the six electron beams at the EPPS slit. Analysis of the scanned image plates together with the determined energy distribution arriving in the spectrometer provides improved dispersion curves for the EPPS.
Related Concept Videos
Mass Spectrometers
Mass Analyzers: Overview
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Atomic Emission Spectroscopy: Instrumentation
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
Atomic Emission Spectroscopy: Lab