Impact of dose reduction and the use of an advanced model-based iterative reconstruction algorithm on spectral
Joël Greffier1, Djamel Dabli1, Aymeric Hamard1
1Department of Medical Imaging, CHU Nimes, Univ Montpellier, Medical Imaging Group Nimes, EA 2992, Bd Pr Robert Debré, 30029 Nîmes Cedex 9, France.
Purpose:
To assess the impact of dose reduction and the use of an advanced modeled iterative reconstruction algorithm (ADMIRE) on image quality in low-energy monochromatic images from a dual-source dual energy computed tomography CT (DSCT) platform.
Materials And Methods:
Acquisitions on an image-quality phantom were performed using DSCT equipment with 100/Sn150 kVp for four dose levels (CTDIvol: 20/11/8/5mGy). Raw data were reconstructed for six energy levels (40/50/60/70/80/100 keV) using filtered back projection and two levels of ADMIRE (A3/A5). Noise power spectrum (NPS) and task-based transfer function (TTF) were calculated on virtual monoenergetic images (VMIs). Detectability index (d') was computed to model the detection task of two enhanced iodine lesions as function of keV.
Results:
Noise-magnitude was significantly reduced between 40 to 70 keV by -56±0% (SD) (range: -56%--55%) with FBP; -56±0% (SD) (-56%--56%) with A3; and -57±1% (SD) (range: -57%--56%) with A5. The average spatial frequency of the NPS peaked at 70 keV and decreased as ADMIRE level increased. TTF values at 50% were greatest at 40 keV and shifted towards lower frequencies as the keV increased. The detectability of both lesions increased with increasing dose level and ADMIRE level. For the simulated lesion with iodine at 2mg/mL, d' values peaked at 70 keV for all reconstruction types, except for A3 at 20 mGy and A5 at 11 and 20 mGy, where d' peaked at 60 keV. For the other simulated lesion, d' values were highest at 40 keV and decreased beyond.
Conclusion:
At low keV on VMIs, this study confirms that iterative reconstruction reduces the noise magnitude, improves the spatial resolution and increases the detectability of enhanced iodine lesions.
More Related Videos
11:38Voluntary Breath-hold Technique for Reducing Heart Dose in Left Breast Radiotherapy
Published on: July 3, 2014
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
Related Concept Videos
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...