Pacemaker-induced ventricular fibrillation during radiofrequency catheter ablation for ventricular tachycardia
Michael C Y Nam1, Nikki Jones1, Simon Claridge1
1Department of Cardiology Royal Bournemouth and Christchurch Hospitals Bournemouth UK.
Abstract:
Prior to ventricular tachycardia ablation, this patient's cardiac implantable electronic device (CIED) was temporarily programmed to backup pacing mode with tachycardia therapies disabled. During radiofrequency energy delivery, the patient developed ventricular fibrillation requiring emergent cardioversion. Electrogram interrogation showed that the CIED switched to noise reversion mode during ablation. The consequent asynchronous pacing resulted in a paced QRS landing on an intrinsic T wave, inducing ventricular fibrillation. This serves as an important reminder that asynchronous pacing consequent to CIED oversensing could occur in any procedure that could cause electromagnetic interference such as radiofrequency cathteter ablation.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
10:46Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
Related Concept Videos
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Dysrhythmias III: Characteristics of Dysrhythmias
Cardiopulmonary Resuscitation III: AED Use
Dysrhythmias II: Classification of Tachyarrhythmias
Dysrhythmias VI: Management of Dysrhythmias