Effects of Simulated COVID-19 Cytokine Storm on Stent Thrombogenicity
Anne Cornelissen1, Matthew Kutyna2, Qi Cheng2
1CVPath Institute, Inc., 19 Firstfield Road, Gaithersburg, MD 20878, United States of America; Department of Internal Medicine I, University Hospital Aachen, RWTH Aachen University, Pauwelsstrasse 30, 52074 Aachen, Germany.
Background:
Cytokine storm-related hypercoagulation may be important in the pathogenesis of stent thrombosis in patients with SARS-CoV-2. Whether stent polymers behave differently under such conditions has never been explored.
Methods:
Fluorinated polymer-nanocoated and uncoated COBRA stents (CeloNova), BioLinx-polymer-coated Resolute Onyx stents (Medtronic), and Synergy stents (Boston Scientific), which are abluminally coated with a bioabsorbable polymer, were exposed to human blood from healthy donors which was supplemented with 400 pg/mL IL-6 and 100 pg/mL TNF-α, similar to what is seen in cytokine storm caused by SARS-CoV-2. Platelet adhesion and neutrophil activation, assessed by immunofluorescence, were compared under cytokine storm and control conditions (untreated blood) (n = 4 experimental runs).
Results:
Platelet adhesion values, defined as %platelet-covered area x staining intensity, were significantly lower in coated and uncoated COBRA and in Resolute Onyx than in Synergy under control conditions (1.28 × 107 ± 0.43 × 107 vs. 2.92 × 107 ± 0.49 × 107 vs. 3.57 × 107 ± 0.73 × 107 vs. 9.94 × 107 ± 0.99 × 107; p ≤0.0001). In cytokine storm, platelet adhesion values remained low in coated COBRA-PzF (1.78 × 107 ± 0.38 × 107) compared to all other devices (uncoated COBRA: 5.92 × 107 ± 0.96 × 107; Resolute Onyx: 7.27 × 107 ± 1.82 × 107; Synergy: 11.28 × 107 ± 1.08 × 107; p ≤ 0.0001). Although cytokine storm conditions significantly increased neutrophil activation in all stents, it was significantly less in coated and uncoated COBRA, and in Resolute Onyx than in Synergy.
Conclusions:
Blood-biomaterials interactions may determine the thrombogenic potential of stents. Under simulated cytokine storm conditions, fluoropolymer-coated stents showed the most favorable anti-thrombogenic and anti-inflammatory properties.
More Related Videos
09:41Hemocompatibility Testing of Blood-Contacting Implants in a Flow Loop Model Mimicking Human Blood Flow
Published on: March 5, 2020
06:37Using Reference Reagents to Confirm Robustness of Cytokine Release Assays for the Prediction of Monoclonal Antibody Safety
Published on: September 15, 2023
Related Concept Videos
Acute Coronary Syndrome I: Introduction
Acute Coronary Syndrome II: Pathophysiology and Clinical Manifestations
Intracellular Signaling Affects Focal Adhesions
Some...