Clustering and Nonclustering Modifier Mixtures in Differential Mobility Spectrometry for Multidimensional Liquid
David Ruskic1, Frank Klont1, Gérard Hopfgartner1
1Life Sciences Mass Spectrometry, Department of Inorganic and Analytical Chemistry, University of Geneva, 24 Quai Ernest Ansermet, CH-1211 Geneva 4, Switzerland.
Abstract:
Modifiers provide fast and reliable tuning of separation in differential mobility spectrometry (DMS). DMS selectivity for separating isomeric molecules depends on the clustering modifier concentration, which is typically 1.5-3 mol % ratio of isopropanol or ethanol in nitrogen. Low concentrations (0.1%) of isopropanol were found to improve resolution and sensitivity but at the cost of practicality and robustness. Replacing the single-channel DMS pump with a binary high-performance liquid chromatography (HPLC) pump enabled the generation of modifier mixtures at a constant flow rate using an isocratic or gradient mode, and the analytical benefits of the system were investigated considering cyclohexane, n-hexane, or n-octane as nonclustering modifiers and isopropanol or ethanol as clustering modifiers. It was found that clustering and nonclustering modifier mixtures enable optimization of selectivity, resolution, and sensitivity for different positional isomers and diastereoisomers. Data further suggested different ion separation mechanisms depending on the modifier ratios. For 85 analytes, the absolute difference in compensation voltages (CoVs) between pure nitrogen and cyclohexane at 1.5 mol % ratio was below 4 V, demonstrating its potential as a nonclustering modifier. Cyclohexane's nonclustering behavior was further supported by molecular modeling using density functional theory (DFT) and calculated cluster binding energies, showing positive ΔG values. The ability to control analyte CoVs by adjusting modifier concentrations in isocratic and gradient modes is beneficial for optimizing multidimensional LCxDMS-MS. It is fast and effective for manipulating the DMS scanning window size to realize shorter mass spectrometry (MS) acquisition cycle times while maintaining a sufficient number of CoV steps and without compromising DMS separation performance.
Related Concept Videos
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Tandem Mass Spectrometry
Secondary fragmentations occur in the interaction cell and can be induced by various factors. Fragmentation induced by collision with inert gases, such as N2, Ar, He, etc., is called collision-induced...
Analyte Adsorption and Distribution
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
High-Resolution Mass Spectrometry (HRMS)
Mass Analyzers: Common Types