Edge area metric complexity scoring of volumetric modulated arc therapy plans
Julia Götstedt1, Anna Bäck1,2
1Department of Radiation Physics, Institute of Clinical Sciences, Sahlgrenska Academy, University of Gothenburg, 413 45 Gothenburg, Sweden.
Background And Purpose:
Aperture-based complexity metrics have been suggested as a method to score complexity of volumetric modulated arc therapy (VMAT) plans. The purpose of this study was to evaluate the edge area metric (EAM) for clinical VMAT plans on a control point and treatment plan level.
Materials And Methods:
EAM on a control point level was evaluated based on film measurements of 18 static beam openings originating from VMAT plans. EAM on a treatment plan level (arithmetic mean value of EAM scores for control points) was evaluated based on measurements with the Delta4® for 200 VMAT plans for four different treatment sites: pelvic, thorax, head and neck, and prostate. Measurements were compared to calculations and dose difference and gamma pass rates were evaluated.
Results:
EAM scores on a control point level correlated with Pearson's r-values of -0.96 and -0.77 to dose difference and gamma pass rates, respectively. The prostate plans had the highest average EAM score. A connection between smaller PTVs and higher EAM scores was found. No correlation between the evaluation result and EAM on a plan level was found.
Conclusions:
EAM on a control point level was shown to correlate to the difference between measured and calculated 2D dose distributions of clinical VMAT beam openings. No correlation was found for EAM on a plan level for clinical treatment plans.
Related Concept Videos
Area Computation by the Alternative Coordinate Method
Accuracy, limits, and approximation
Accuracy is defined as the closeness of the measured value to the true or actual value. In engineering mechanics, repeated measurements are taken during theoretical or experimental analyses to ensure that the result is precise and accurate.
The accuracy of any solution is based on the...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Areas Within Irregular Boundaries