Factors Affecting the Robustness of Data Inversion for Stable Isotope Measurement Using the Double Spike Method:
Xiangli Wang1,2, Thomas M Johnson3
1University of South Alabama, Mobile, Alabama 36688, United States.
Abstract:
Stable isotope ratios are widely used to solve environmental, geological, medical, and forensic problems. The double spike technique is considered to be one of the most robust and efficient methods to correct for instrumental mass bias and isotopic fractionation that may occur during sample preparation. However, various hidden errors can arise from data processing and have been largely overlooked in previous studies. Several of these hidden errors were investigated in this work using measurement and synthetic data. Double spike inversion of chromium isotope raw data from 1116 natural samples demonstrated that averaging raw isotope ratios before double spike inversion can add significant errors to inverted isotope values, and such errors can be 1.5 times larger than the true analytical precision. Synthetic data were used to investigate the errors on inverted Cr isotope data caused by spike:analyte ratio and Fe-Ti-V interferences, and the following threshold values are recommended to minimize such errors: 54Crspike/52Crsample ratio greater than 0.5, 56Fe/52Cr less than 0.2, 49Ti/52Cr less than 0.04, and 51V/52Cr less than 1. Sample preparation can potentially lead to large errors in inverted Cr isotope data if preparation-induced isotope fractionation deviates from the exponential law used in the double spike inversion, but such errors can be minimized by achieving >70% Cr yield. Our findings provide important insights for the double spike inversion procedure and assessing the reliability of inverted isotope data for not only the chromium isotope system but also other elements commonly analyzed using the double spike technique.
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Voltammetry: Factors Affecting Measurements
Mass Spectrometry: Isotope Effect
¹³C NMR: ¹H–¹³C Decoupling
A broadband decoupling technique is used to simplify these complex, sometimes overlapping, signals. Broadband decoupling relies on a...
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Nuclear Stability
To hold positively charged protons together...