Pediatric reference interval verification for common biochemical assays on the Abbott Alinity system.
Mary Kathryn Bohn1,2, Siobhan Wilson1,2, Alexandra Hall1
1CALIPER Program, Pediatric Laboratory Medicine, The Hospital for Sick Children, Toronto, ON, Canada.
Clinical Chemistry and Laboratory Medicine
|May 19, 2021
Summary
Pediatric reference intervals (RIs) from the CALIPER initiative for 28 chemistry assays on the Abbott ARCHITECT were verified on the Alinity system. Most assays showed concordance, aiding the Alinity system
Area of Science:
- Clinical Chemistry
- Laboratory Medicine
- Pediatric Reference Intervals
Background:
- Accurate clinical laboratory service relies on validated reference intervals (RIs).
- New analytical systems necessitate verification of established RIs.
- The Canadian Laboratory Initiative on Pediatric Reference Intervals (CALIPER) provides RIs for pediatric populations.
Purpose of the Study:
- To verify published CALIPER pediatric RIs for 28 chemistry assays on the Abbott ARCHITECT system using the newer Alinity system.
- To assess the concordance between Abbott ARCHITECT and Alinity analytical systems for pediatric chemistry assays.
Main Methods:
- Analytical validation of 28 chemistry assays on the Alinity system using 100 CALIPER serum samples.
- Comparison of Alinity results against published pediatric RIs and confidence limits from the Abbott ARCHITECT system.
- Verification criteria based on Clinical and Laboratory Standards Institute (CLSI) guidelines (≥90% of results within confidence limits).
Main Results:
- 26 out of 28 assays met the verification criteria.
- Calcium (87%) and magnesium (35%) assays did not meet verification criteria based on ARCHITECT RIs.
- Calcium and magnesium assays were verifiable using Abbott Alinity package insert RIs.
Conclusions:
- CALIPER ARCHITECT RIs were successfully verified on the Alinity system for most assessed chemistry assays.
- Demonstrated concordance between Abbott ARCHITECT and Alinity systems supports the implementation of the Alinity system in pediatric settings.
- Verification of RIs is crucial for accurate interpretation of pediatric laboratory test results.