Wireless Bladder Pressure Monitor for Closed-Loop Bladder Neuromodulation
Steve Majerus1,2, Anisha S Basu3,4, Iryna Makovey5
1Advanced Platform Technology Center, Cleveland VA Medical Center, Cleveland, OH, USA.
Summary
A new wireless bladder pressure monitor offers catheter-free, real-time detection for conditional neuromodulation. Chronic implantation in large animals confirmed device function and data quality for bladder control.
Area of Science:
- Biomedical Engineering
- Urology
- Implantable Devices
Background:
- Conditional neuromodulation requires real-time bladder pressure monitoring.
- Current catheter-based methods are unsuitable for chronic ambulatory therapy.
- A wireless, implantable solution is needed for closed-loop bladder control.
Purpose of the Study:
- To develop and evaluate a wireless, implantable pressure monitor for bladder control.
- To assess the viability of a non-hermetic polymer sensor package for chronic implantation.
- To confirm chronic wireless device function and data quality in large animal models.
Main Methods:
- Developed an ultra-low-power ASIC, MEMS pressure sensor, RF antennas, and rechargeable battery system.
- Utilized a non-hermetic polymer package for sensor encapsulation.
- Conducted chronic in vivo testing in large animal models.
Main Results:
- The polymer packaging method proved viable for chronic sensor encapsulation.
- Chronic implantation presented some challenges at the bladder wall site.
- Confirmed chronic wireless device function with high data quality (avg. correlation coefficient of 0.90) for detecting bladder contractions.
Conclusions:
- The wireless pressure monitor is a viable catheter-free solution for real-time bladder pressure monitoring.
- The developed system demonstrates potential for chronic ambulatory therapy in bladder control.
- Further optimization of the implant site is recommended for long-term stability.