A novel CAD/CAM composite occlusal splint for intraoperative verification in single-splint two-jaw orthognathic
Lun-Jou Lo1, Lien-Shin Niu2, Chun-Hao Liao2
1Department of Plastic and Reconstructive Surgery, and Craniofacial Research Center, Chang Gung Memorial Hospital at Linkou, Taoyuan, Taiwan; College of Medicine, Chang Gung University, Taoyuan, Taiwan.
Background:
Previous computer-generated splints were designed and produced without modification than the traditional occlusal splints, which did not facilitate surgeon's intraoperative judgment in the single-splint two-jaw orthognathic surgery. Modifications of the digital occlusal splint can be achieved using computer-aided design and computer-aided manufacturing (CAD/CAM) software. This study reported the design, clinical application and validation of a novel CAD/CAM occlusal splint.
Methods:
The maxillary and mandibular segments were fixed into the final occlusal splint and moved to the planned position according to the 3-dimensional simulation. The composite occlusal splint has 4 orthogonal bars to facilitate intraoperative assessment of the dental and skeletal midline, facial soft tissue midline, occlusal plane, upper tooth show, facial symmetry and facial bone position. To validate the surgical outcome, 5 parameters including pitch, roll and yaw rotations, midline deviation and chin position were measured on the virtual plan and the postoperative cone-beam computed tomography images to quantify the difference.
Results:
The results showed no significant differences in the 5 parameters between the simulation and postoperative images. The root-mean-square difference between the conventional splints and CAD/CAM surgical splint ranged from 0.18 to 0.31 mm by superimposition of the two image models. All patients were satisfied with the treatment outcomes. Overall, this novel occlusal splint is ideal for verification of the maxillomandibular position during surgery.
Conclusion:
The novel composite occlusal splint provided useful and informative check to verify the maxillomandibular complex (MMC) position and facial appearance in single-splint two-jaw orthognathic surgery.