Related Experiment Video
Updated: Nov 1, 2025

Laryngeal Mask Airway LMA Placement in a Neonatal Patient Simulator Using a Non-Inflatable Supraglottic Airway SGA
Published on: July 14, 2023
Assisted ventilation immediately after birth with self-inflating bag versus T-piece resuscitator in preterm infants
Objective:
To compare proportions of target range tidal volumes achieved with the self-inflating bag vs. the T-piece in resuscitation of preterm newborns at delivery.
Study Design:
This randomized controlled trial was conducted at a tertiary Children's Hospital. 20 preterm infants≤32 weeks' gestational age with no congenital anomalies who needed positive pressure ventilation after birth were enrolled. Positive pressure ventilation was provided with the self-inflating bag or T-piece resuscitator. The primary outcome was proportion of inflations within a target range of 4-8 ml/kg. Chi-square and logistical regression analyses were performed.
Results:
In the self-inflating bag (SIB) group 29% of inflations (117/419) and in the T-Piece (TP) group 51% of inflations (300/590) delivered expiratory tidal volume (TVe) of 4-8 ml/kg (p < 65.001). In the SIB group 60% of all inflations (254/419), and in the TP group 35% of all inflations (204/590) delivered TVe < 4 ml/kg (p < 0.001). In the SIB group 11% of all inflations (48/419), and in the TP group, 15% of all inflations (86/590) delivered TVe > 8 ml/kg (p = 0.18). The OR of having expiratory tidal volume of 4-8 ml/kg using the T-piece was 1.8 (CI 1.1-3.1), p = 0.02.
Conclusion:
Manual inflations provided by the TP deliver expiratory tidal volumes in the range of 4-8 ml/kg more consistently than SIB.
Related Concept Videos
Cardiopulmonary Resuscitation II: ACLS Airway Management
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...