Noise characterization and compensation for a charge-coupled-device-based pyrometer
Yuzhong Zhang1, Fucheng Lu1, Wenjing Wang1
1School of Instrument Science and Opto-Electronics Engineering, Hefei University of Technology, Hefei 230009, Anhui, China.
Abstract:
The pyrometer based on the charge coupled device (CCD) is a cost-effective and widely used system for temperature measurement in various industrial fields. However, due to the inter-element sensitivity deviations of the CCD-array detector and the influence of various types of noise, the digital signal output of the CCD sensor itself is not completely equal to the ideal value as expected. In this work, based on a classical calibration method, the inter-element sensitivity deviations, dark current, shot noise, and readout-quantization noise for a CCD-based pyrometer are characterized, and the influence of these noises on the temperature measurement accuracy is evaluated quantitatively. Furthermore, the non-uniformity correction coefficient for each pixel is obtained by a simple segment correction method to reduce the inter-element sensitivity deviations, and meanwhile, the Kalman filter is introduced to remove the temperature fluctuations caused by these noises. Experimental results show that mean value of the spatial standard deviation of temperature measurement results in the total measurement range (800-1200 °C) is only 2.2 °C after non-uniformity correction, and the temperature fluctuations can be reduced from 26.6 to 1.98 °C based on the Kalman filter.
More Related Videos
07:44Characterization of Full Set Material Constants and Their Temperature Dependence for Piezoelectric Materials Using Resonant Ultrasound Spectroscopy
Published on: April 27, 2016
10:22Split Point Analysis and Uncertainty Quantification of Thermal-Optical Organic/Elemental Carbon Measurements
Published on: September 7, 2019
Related Concept Videos
Gas Chromatography: Types of Detectors-II
Flame Photometry: Overview
Flame Photometry: Lab
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...