Falsely diagnosed thyrotoxicosis caused by anti-streptavidin antibodies and pre-wash procedures
Takuya Ishikawa1, Hiroyuki Sakai2, Tokutaro Itaya1
1Department of Diabetes, Metabolism and Endocrinology, Tokyo Medical University, 6-7-1 Nishishinjuku Shinjuku-ku, Tokyo, 160-0023, Japan.
Background:
Anti-streptavidin antibodies are causal determinants of analytical interference during Thyroid function tests, and numerous reports have detailed such interference, with anti-streptavidin antibodies attracting attention.
Case Presentation:
We conducted a straightforward investigation of interference due to anti-streptavidin antibodies, with a case of a 60-year-old Japanese man who consulted our department for inconsistencies between his clinical course and Thyroid function tests. Experiments were conducted using Cobas8000 e602, which employs assay procedures with pre-wash to evaluate FT4 and FT3 levels.
Conclusions:
To our knowledge, this is the first published report to clearly investigate such interferences using a combination of polyethylene glycol precipitation, heterophilic blocking tube precipitation, streptavidin-coated magnetic particle precipitation, and different instruments with or without pre-wash. Clinicians should consider that interferences caused by anti-streptavidin antibodies could lead to a misdiagnosis of thyrotoxicosis. Moreover, discussions between laboratory specialists, clinicians, and manufacturers are required to identify interferences and avoid unnecessary examinations and inappropriate treatment.