The accuracy of heartbeat detection using photoplethysmography technology in cardiac patients
S Blok1, M A Piek2, I I Tulevski2
1Department of Cardiology, Cardiology Centers of the Netherlands, Karel du Jardinstraat 61, Amsterdam, the Netherlands; Department of Vascular Medicine, Internal Medicine, Amsterdam UMC, University of Amsterdam, Amsterdam Cardiovascular Sciences, Meibergdreef 9, Amsterdam, the Netherlands.
Introduction:
Photoplethysmography (PPG) in wearable sensors potentially plays an important role in accessible heart rhythm monitoring. We investigated the accuracy of a state-of-the-art bracelet (Corsano 287) for heartbeat detection in cardiac patients and evaluated the efficacy of a signal qualifier in identifying medically useful signals.
Methods:
Patients from an outpatient cardiology clinic underwent a simultaneous resting ECG and PPG recording, which we compared to determine accuracy of the PPG sensor for detecting heartbeats within 100 and 50 ms of the ECG-detected heart beats and correlation and Limits of Agreement for heartrate (HR) and RR-intervals. We defined subgroups for skin type, hair density, age, BMI and gender and applied a previously described signal qualifier.
Results:
In 180 patients 7914 ECG-, and 7880 (99%) PPG-heartbeats were recorded. The PPG-accuracy within 100 ms was 94.6% (95% CI 94.1-95.1) and 89.2% (95% CI 88.5-89.9) within 50 ms. Correlation was high for HR (R = 0.991 (95% CI 0.988-0.993), n = 180) and RR-intervals (R = 0.891 (95% CI 0.886-0.895), n = 7880). The 95% Limits of Agreement (LoA) were -3.89 to 3.77 (mean bias 0.06) beats per minute for HR and -173 to 171 (mean bias -1) for RR-intervals. Results were comparable across all subgroups. The signal qualifier led to a higher accuracy in a 100 ms range (98.2% (95% CI 97.9-98.5)) (n = 143).
Conclusion:
We showed that the Corsano 287 Bracelet with PPG-technology can determine HR and RR-intervals with high accuracy in cardiovascular at-risk patient population among different subgroups, especially with a signal quality indicator.
More Related Videos
05:03Patient Directed Recording of a Bipolar Three-Lead Electrocardiogram using a Smartwatch with ECG Function
Published on: December 11, 2019
05:51Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Special considerations while measuring pulse
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Holter Monitor: 24-Hour Monitoring
Errors occurring during blood pressure monitoring
Several factors...