Measuring the Carotid to Femoral Pulse Wave Velocity Cf-PWV to Evaluate Arterial Stiffness
Published on: May 3, 2018
Blood pressure measurement: Should technique define targets?
Swapnil Hiremath1,2,3, Tim Ramsay2, Marcel Ruzicka1,3
1Department of Medicine, University of Ottawa, Ottawa, Ontario, Canada.
Medical personnel presence does not affect automated blood pressure (BP) readings. Resting office BP measurements underestimate daytime BP load compared to 24-h ambulatory BP monitoring.
Area of Science:
- Cardiology
- Hypertension Management
- Medical Device Technology
Background:
- Accurate blood pressure (BP) assessment is vital for hypertension management.
- Automated oscillometric BP (AOBP) devices are commonly used for BP measurement.
- The impact of medical personnel presence on AOBP accuracy requires clarification.
Purpose of the Study:
- To quantify the effect of medical personnel presence on AOBP measurements.
- To compare resting office BP by AOBP with 24-h ambulatory BP monitoring (ABPM).
Main Methods:
- Prospective randomized cross-over trial in a referral population.
- Patients underwent casual and resting office BP measurements using AOBP (Omron HEM 907XL).
- Resting BP was measured unattended and partially attended, before and after 24-h ABPM.
Main Results:
- Casual office BP was significantly higher than unattended resting office BP (mean difference 7.0 mm Hg).
- No significant difference in resting office systolic BP between partially attended and unattended measurements.
- Resting office BP underestimated daytime systolic BP load from 24-h ABPM.
Conclusions:
- Medical personnel presence does not impact casual office BP readings.
- The requirement for unattended rest before AOBP may be relaxed if logistically difficult.
- Office BP measurements by AOBP do not fully represent the complexity of 24-h ABPM data.
Related Concept Videos
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Errors occurring during blood pressure monitoring
Several factors...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Blood Pressure
The average BP in an adult is typically around 120/80 mmHg (millimeters of mercury). In this measurement, the numerator (120) indicates the systolic pressure, which is the pressure in the arteries during the contraction of the heart's ventricles as blood is expelled. The denominator (80) represents the...