Determining EMC Test Levels for Implantable Devices in Bipolar Lead Configuration.
Biomedical Instrumentation & Technology
|July 19, 2021
Summary
Electromagnetic compatibility testing for pacemakers and implantable cardioverter defibrillators (ICDs) uses lower test levels for bipolar leads. This study confirms the one-tenth amplitude reduction for bipolar configurations is clinically relevant and should continue.
Area of Science:
- Biomedical Engineering
- Electromagnetics
- Medical Device Standards
Background:
- Low-frequency magnetic fields can interfere with implantable medical devices like pacemakers and implantable cardioverter defibrillators (ICDs).
- Electromagnetic compatibility (EMC) standards simplify testing by simulating real-world exposure, but the rationale for specific test levels is crucial.
- The ISO 14117:2012 standard specifies a 10:1 reduction in test levels for bipolar lead configurations compared to unipolar configurations for pacemakers and ICDs.
Purpose of the Study:
- To investigate the clinical relevance and provide a scientific basis for the reduced EMC test levels specified for bipolar lead configurations in pacemakers and ICDs.
- To evaluate the difference in magnetically induced voltages between unipolar and bipolar lead configurations under simulated low-frequency magnetic field exposure.
Main Methods:
- Analytical calculations using Faraday's law to determine magnetically induced voltages in a 37.6-cm lead with varying bipolar electrode spacing.
- Computational modeling using electromagnetic field simulation software to analyze induced voltages in different lead configurations immersed in a saline-conducting medium.
- Comparison of induced voltages between bipolar and unipolar lead configurations under parallel, closely spaced conditions.
Main Results:
- Both analytical calculations and computational modeling demonstrated that induced voltages in bipolar lead configurations are significantly lower than in unipolar configurations.
- The results support the existing practice of using a one-tenth amplitude reduction for testing bipolar configurations of pacemakers and ICDs.
- The study provides a quantitative basis for the EMC test level difference specified in the ISO 14117 standard.
Conclusions:
- The findings validate the continued use of the one-tenth amplitude reduction for EMC testing of pacemakers and ICDs in bipolar mode.
- This study's rationale has been incorporated into the most recent edition of the ISO 14117 standard, ensuring accurate and relevant device testing.
- The research contributes to the safety and reliability of implantable medical devices by refining electromagnetic compatibility testing protocols.