Measuring radiofrequency fields in NMR spectroscopy using offset-dependent nutation profiles

Ahallya Jaladeep1, Claris Niya Varghese1, Ashok Sekhar1

  • 1Molecular Biophysics Unit, Indian Institute of Science, Bangalore 560 012, India.

Summary

Accurately measure radiofrequency (RF) field strengths in NMR spectroscopy using intrinsic molecular modulations. This method precisely determines B1 field values and distributions, even for large biomolecules with external standards.

Related Concept Videos

NMR Spectrometers: Overview01:20

NMR Spectrometers: Overview

NMR spectrometers consist of a strong magnet, a radiofrequency transmitter, and a detector attached to a computer console for recording spectra of samples containing NMR-active nuclei. In first-generation NMR instruments called continuous-wave spectrometers, the resonance frequencies of the nuclei are determined by frequency-sweep or field-sweep methods. The magnetic field strength is fixed and the rf signal is swept in the former, while the radiofrequency signal is fixed and the magnetic field...
1.5K
¹H NMR: Interpreting Distorted and Overlapping Signals01:02

¹H NMR: Interpreting Distorted and Overlapping Signals

Spin systems where the difference in chemical shifts of the coupled nuclei is greater than ten times J are called first-order spin systems. These nuclei are weakly coupled, and their chemical shifts and coupling constant can generally be estimated from the well-separated signals in the spectrum.
As Δν decreases and the signals move closer, the doublets appear increasingly distorted. The intensities of the inner lines increase at the cost of those of the outer lines as the signals are...
1.2K
NMR Spectrometers: Resolution and Error Correction01:14

NMR Spectrometers: Resolution and Error Correction

When magnetic nuclei in a sample achieve resonance and undergo relaxation, the signal detected in NMR is an approximately exponential free induction decay. Fourier transform of an exponential decay yields a Lorentzian peak in the frequency domain. Lorentzian peaks in an NMR spectrum are defined by their amplitude, full width at half maximum, and position, where the peak width is governed by the spin-spin relaxation time alone. In real experiments, however, the applied magnetic field is rendered...
826
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences01:17

NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences

A pulse is a short burst of radio waves distributed over a range of frequencies that simultaneously excites all the nuclei in the sample. Upon passing a radio frequency pulse along the x-axis, the nuclei absorb energy corresponding to their Larmor frequencies and achieve resonance. This shifts the net magnetization vector from the z-axis toward the transverse plane. This angle of rotation of the magnetization vector, or the flip angle, is proportional to the duration and intensity of the pulse.
1.0K
¹H NMR of Conformationally Flexible Molecules: Variable-Temperature NMR01:15

¹H NMR of Conformationally Flexible Molecules: Variable-Temperature NMR

The axial and equatorial protons in cyclohexane can be distinguished by performing a variable-temperature NMR experiment. In this process, except for one proton, the remaining eleven protons are replaced by deuterium. The deuterium substitution avoids the possible peak splitting caused by the spin-spin coupling between the adjacent protons. The remaining proton flips between the axial and equatorial positions.
1.3K
Double Resonance Techniques: Overview01:12

Double Resonance Techniques: Overview

Double resonance techniques in Nuclear Magnetic Resonance (NMR) spectroscopy involve the simultaneous application of two different frequencies or radiofrequency pulses to manipulate and observe two distinct nuclear spins. One important application of double resonance is spin decoupling, which selectively suppresses coupling with one type of nucleus while observing the NMR signal from another nucleus, simplifying the spectrum and enhancing resolution.
Spin decoupling is usually achieved by...
376