Monitoring of carbon dioxide in ventilated neonates: a prospective observational study
Tobias Werther1, Lukas Aichhorn2, Sina Stellberg2
1Department of Pediatrics, Division of Neonatology, Pediatric Intensive Care and Neuropediatrics, Comprehensive Center for Pediatrics, Medical University of Vienna, Vienna, Austria tobias.werther@meduniwien.ac.at.
Objective:
To assess the reliability, accuracy and precision of distal end-tidal capnography (detCO2) in neonates compared with transcutaneous (tcCO2) carbon dioxide measurements.
Design:
Observational, prospective clinical study.
Setting:
Neonatal intensive care unit at Medical University of Vienna.
Participants:
Conventionally ventilated neonates with a body weight between 1000 g and 3000 g.
Intervention:
End-tidal partial pressure of CO2 was measured in distal position using the separate lumen of a double-lumen endotracheal tube connected to an external side-stream capnometer. Three consecutive detCO2 and tcCO2 values were recorded simultaneously and compared with simultaneous arterialised partial pressure of CO2 (paCO2) measurements in each patient.
Main Outcome Measures:
Reliability, accuracy and precision of detCO2 and tcCO2 measurements compared with paCO2 in neonates.
Results:
Twenty-five neonates were included with a median (range) weight at enrolment of 1410 (1010-2980) g, from which 81 simultaneous measurements of detCO2, tcCO2 and paCO2 were obtained. The mean (SD) of paCO2, detCO2 and tcCO2 was 45.0 (8.6) mmHg, 42.4 (8.4) mmHg and 50.4 (20.4) mmHg, respectively. The intraclass correlation between paCO2 and detCO2 and between paCO2 and tcCO2 reached 0.80 (95% CI 0.71 to 0.87, p<0.001) and 0.59 (95% CI 0.43 to 0.72, p<0.001), respectively. In the Bland-Altman analysis, bias and precision of detCO2 with respect to paCO2 amounted to -2.68 mmHg and 10.62 mmHg (95% CI 8.49 to 14.51), respectively. Bias and precision of tcCO2 with respect to paCO2 amounted to 5.39 mmHg and 17.22 mmHg (95% CI 13.21 to 23.34), respectively.
Conclusion:
DetCO2 had better reliability, accuracy and precision with paCO2 than tcCO2 in ventilated neonates without severe lung diseas.
Trial Registration Number:
NCT03758313.
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Respiratory Regulation of Acid-Base Balance
When carbon dioxide levels increase in the blood, more H+ and HCO3⁻ are...
Assessment of Respiration
Subjective Assessment: Nurses interview the patient to gather information directly during the subjective assessment. It includes questions about the individual's medical history, medications, and symptoms, focusing on past respiratory conditions like...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...