Nix Pro Color Sensor provides comparable color measurements to HunterLab colorimeter for fresh beef
Conrad S Schelkopf1, Emily A Rice1, Joanna K Swenson1
1Center for Meat Safety & Quality, Department of Animal Sciences, Colorado State University, 1171 Campus Delivery, Fort Collins, CO 80523-1171 USA.
Abstract:
The HunterLab MiniScan (HunterLab) colorimeter is used in meat quality research worldwide for measuring meat color; however, the Nix Pro Color Sensor (Nix) could be a less expensive alternative that is easier to operate. Therefore, the objective of this study was to compare the two colorimeters to objectively evaluate fresh beef color. Longissimus thoracis muscle from one side of A maturity beef carcasses (n = 200) was evaluated using both the HunterLab (3 technical replicate scans) and Nix (3, 5, 7, and 9 technical replicate scans) colorimeters. The correlation between the HunterLab and Nix for L* (lightness), a* (redness), and b* (yellowness) values ranged between r = 0.80 to 0.85 and the Bland Altman Limits of Agreement analysis indicated good agreement between the Nix and HunterLab colorimeters for all the color parameters. These results indicated that the Nix colorimeter could be a viable alternative for HunterLab colorimeters.