Low-dose CT denoising via convolutional neural network with an observer loss function

Minah Han1, Hyunjung Shim1, Jongduk Baek1

  • 1School of Integrated Technology and Yonsei Institute of Convergence Technology, Yonsei University, Incheon, South Korea.

Medical Physics
|August 13, 2021
PubMed
Abstract

Related Concept Videos

Reducing Line Loss01:18

Reducing Line Loss

In a three-phase circuit, line loss is an indicator of energy dissipated as heat due to the resistance of transmission lines. To address this, incorporating transformers into the system—a step-up transformer at the source and a step-down transformer at the load—is a strategic solution. Two three-phase transformers are introduced to improve this.
With a step-up transformer at the source, the voltage is increased, thereby reducing the current in the transmission lines since power loss...
224
Deconvolution01:20

Deconvolution

Deconvolution, also known as inverse filtering, is the process of extracting the impulse response from known input and output signals. This technique is vital in scenarios where the system's characteristics are unknown, and they must be inferred from the observable signals.
Deconvolution involves several mathematical techniques to derive the impulse response. One common approach is polynomial division. In this method, the input and output sequences are treated as coefficients of...
335
Computed Tomography01:10

Computed Tomography

Tomography refers to imaging by sections. Computed tomography (CT) is a non-invasive imaging technique that uses computers to analyze several cross-sectional X-rays to reveal minute details about structures in the body.
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
7.3K
Downsampling01:20

Downsampling

When considering a sampled sequence with zero values between sampling instants, one can replace it by taking every N-th value of the sequence. At these integer multiples of N, the original and sampled sequences coincide. This process, known as decimation, involves extracting every N-th sample from a sequence, thereby creating a more efficient sequence.
The Fourier transform of the decimated sequence reveals a combination of scaled and shifted versions of the original spectrum. This...
329
Imaging Studies III: Computed Tomography01:27

Imaging Studies III: Computed Tomography

DefinitionComputed Tomography (CT) of the genitourinary (GU) tract is a non-invasive imaging modality that utilizes X-rays and computer processing to generate detailed cross-sectional images of the urinary system, encompassing the kidneys, ureters, bladder, and adjacent structures such as the adrenal glands.PurposeCT scans of the GU tract serve several diagnostic and therapeutic purposes, including:Diagnosis of Urinary Tract Diseases: Detects kidney stones, tumors, cysts, and congenital...
94
Difference from Background: Limit of Detection01:05

Difference from Background: Limit of Detection

The limit of detection (LOD) is the smallest amount of analyte that can be distinguished from the background noise. The LOD value corresponds to the concentration at which the analyte signal is three times larger than the standard deviation of the blank signal. Below this value, the analyte signal cannot be differentiated from the background noise. It is calculated by dividing the calibration slope by 3 times the standard deviation of the blank signals.
The LOD indicates the presence or absence...
7.5K