kV-kV and kV-MV DECT based estimation of proton stopping power ratio - a simulation study
Kai-Wen Li1, Daiyu Fujiwara2, Akihiro Haga2
1Beijing Advanced Innovation Center for Big Data-Based Precision Medicine, School of Medicine and Engineering, Beihang University, Key Laboratory of Big Data-Based Precision Medicine (Beihang University), Ministry of Industry and Information Technology, Beijing 100191, China; School of Physics, Beihang University, Beijing 102206, China.
Purpose:
This study aims to estimate the proton stopping power ratio (SPR) by using 80-120 kV and 120 kV-6 MV dual-energy CT (DECT) in a fully simulation-based approach for proton therapy dose calculations.
Methods:
Based on a virtual CT system, a two-step approach is applied to obtain the reference attenuation coefficient for image reconstruction. The effective atomic number (EAN) and electron density ratio (EDR) are estimated from two CT scans. The SPR is estimated using a calibration based on known materials to obtain a piecewise linear relationship between the EAN and the logarithm of the mean excitation energy, lnIm. The calibration phantoms are constructed from reference tissue materials taken from ICRU Report 44. Our approach is evaluated through using the ICRP110 human phantom. The respective influences of noise and beam hardening effects are studied.
Results:
With the beam hardening correction applied, the results of 120 kV-6 MV DECT are comparable to those of 80-120 kV DECT in predicting the EAN, but the former demonstrated a clear improvement in predicting the EDR and SPR. The 120 kV-6 MV DECT is able to predict the SPR within an accuracy of 10% for lung tissue and 5% for pelvis tissue, thereby outperforming the 80-120 kV DECT.
Conclusions:
The 120 kV-6 MV DECT is less sensitive to noise but more susceptible to beam hardening effects. By applying beam hardening correction, the 120 kV-6 MV DECT can predict the SPR more accurately than the 80-120 kV DECT. To utilize our DECT approach most effectively, high-quality reconstructed images are required.
More Related Videos
08:31Sample Preparation and Experimental Design for In Situ Multi-Beam Transmission Electron Microscopy Irradiation Experiments
Published on: June 27, 2022
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Related Concept Videos
Estimation of k and VD of Aminoglycosides
¹³C NMR: Distortionless Enhancement by Polarization Transfer (DEPT)
Atomic Nuclei: Nuclear Spin State Population Distribution
Finding Electric Potential From Electric Field