Ideal Illumination for Smartphone-based Trabeculectomy Bleb Photography
Gagan Kalra1, Parul Ichhpujani1, Sahil Thakur2
1Department of Ophthalmology, Government Medical College and Hospital, Sector-32, Chandigarh, India.
Smartphone photography reliably assesses trabeculectomy blebs. Optimal intermediate or distant illumination with white or yellow light minimizes digital noise and enhances morphological details for better clinical documentation.
Area of Science:
- Ophthalmology
- Medical Imaging
- Glaucoma Surgery
Background:
- Smartphones offer novel applications in ophthalmology, including fundus photography and telemedicine.
- Anterior segment photography for trabeculectomy bleb morphology assessment using smartphones remains underexplored.
Purpose of the Study:
- To characterize ideal illumination for anterior segment smartphone photography in patients post-trabeculectomy.
- To evaluate the reliability of smartphone anterior segment photography for bleb morphology assessment.
Main Methods:
- Thirty patients post-trabeculectomy were enrolled.
- Images were captured using native and FiLMiC Pro apps on iPhone X with yellow and white pen-torches.
- ImageJ software and R software were used for area measurement, consistency analysis (ANOVA), and repeatability (Bland-Altman plots).
- Signal-to-noise ratio (SNR) and inter-observer agreement (kappa coefficient) were calculated.
Main Results:
- No significant difference in measured bleb area between the two camera apps, indicating consistency.
- Bland-Altman analysis confirmed good repeatability and reproducibility of measurements.
- Lower SNR was observed with close illumination compared to intermediate and distant illumination.
- Substantial agreement between observers was found for images captured with distant white light (kappa=0.7) and intermediate yellow light (kappa=0.65).
Conclusions:
- Smartphone photography is a reliable tool for assessing trabeculectomy bleb morphology.
- Optimal intermediate or distant illumination is crucial for high SNR, minimizing digital noise and overexposure.
- Proper illumination enhances the delineation of morphological features, improving diagnostic accuracy.