Double-vertebral segment SBRT via novel ring-mounted Halcyon Linac: Plan quality, delivery efficiency and accuracy
Damodar Pokhrel1, Joseph Stephen1, Aaron Webster1
1University of Kentucky, Department of Radiation Medicine, Lexington, KY, USA.
Summary
Single-isocenter/multi-target VMAT for double-vertebral SBRT on Halcyon RDS is feasible. Halcyon plans offer superior OAR sparing and faster delivery compared to Truebeam, improving patient outcomes and workflow.
Area of Science:
- Radiation Oncology
- Medical Physics
- Radiotherapy Technology
Background:
- Stereotactic body radiation therapy (SBRT) for spinal targets requires precise dose delivery.
- Single-isocenter/multi-target volumetric modulated arc therapy (SIMT VMAT) is an advanced technique for treating multiple targets simultaneously.
- The Halcyon ring delivery system (RDS) offers potential advantages in speed and accuracy for complex radiotherapy.
Purpose of the Study:
- To evaluate the plan quality, treatment delivery efficiency, and accuracy of SIMT VMAT for double-vertebral segments SBRT on the Halcyon RDS.
- To compare Halcyon RDS plans with those generated on an SBRT-dedicated Truebeam system.
Main Methods:
- Six patient cases with double-vertebral segments were replanned on Halcyon RDS using 6MV-FFF and SIMT VMAT.
- Plans were compared to retrospective Truebeam plans using identical geometry and objectives, evaluated per NRG-BR002 protocol.
- Phantom testing and independent dose verification were performed; treatment delivery parameters and portal dosimetry QA were assessed.
Main Results:
- Both Halcyon and Truebeam plans met NRG-BR002 compliance criteria with comparable target coverage and dose indices.
- Halcyon plans demonstrated significantly lower maximal and volumetric doses to organs-at-risk (OARs), including the spinal cord.
- Halcyon plans showed reduced monitor units, modulation, and overall treatment time, with higher portal dosimetry QA pass rates.
Conclusions:
- SIMT VMAT for double-vertebral segments SBRT on Halcyon RDS is feasible and dosimetrically superior to Truebeam.
- The Halcyon system provides more accurate treatment delivery and improved OAR sparing.
- Faster treatment delivery on Halcyon can enhance patient comfort, reduce motion artifacts, and improve clinical workflow efficiency.