High Resolution 3D Imaging of Ex-Vivo Biological Samples by Micro CT
Published on: June 21, 2011
3D-printed multisampling holder for microcomputed tomography applied to life and materials science research
Isabel Vasconcelos1, Margarida Franco2, Mário Pereira1
1Department of Endodontics, Faculdade de Medicina Dentária, Universidade de Lisboa, Rua Professora Teresa Ambrósio, Cidade Universitária, 1600-277, Lisboa, Portugal; Center for Rapid and Sustainable Product Development, Polytechnic Institute of Leiria, Rua de Portugal, 2430-028, Marinha Grande, Portugal.
Abstract:
The aim of this work was to design, fabricate, test and validate a 3D-printed multisampling holder for multi-analysis by microcomputed tomography. Different raw materials were scanned by microcomputed tomography. The raw material chosen was used to fabricate the holder by 3D printing. To validate the multisampling holder, five teeth were filled with a high density-material and scanned in two ways: a single and a multisampling scan mode. For each tooth, the root canal filling volume, porosity volume, closed pore volume, and open pore volume were calculated and compared when the same tooth was scanned in the two sampling scan mode. ABSplus P430™ allowed a high transmission value (84.3 %), and then it was the polymeric material selected to fabricate the holder. In a single sampling scan mode, the scan duration for scanning five teeth was 87.42 min, contrasting with 21.51 min for a multisampling scan mode, which scanned five teeth at the same time. The scan duration time and the cost using a multisampling holder represented a reduction of 75 % and the data volume generated represented a reduction of 60 %. Comparing the two scan modes, the results also showed that the difference of root canal filling volume, porosity volume, closed pore volume, and open pore volume was not statistically significant (p > .05). The multisampling holder was validated to do multi-analysis by microcomputed tomography without significant loss of quantitative accuracy data, allowing a reduction in scan duration time, imaging cost, and data storage.
Related Concept Videos
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Three-Dimensional Microscopy in Microbiology