Interpretation of intrafraction motion review data and method for verification
Gavin Graeper1, Ashley Cetnar1, Ahmet S Ayan1
1Department of Radiation Oncology, The Ohio State University Wexner Medical Center, Columbus, Ohio, USA.
Journal of Applied Clinical Medical Physics
|September 28, 2021
Summary
This study introduces a method to extract quantitative intrafraction motion review (IMR) data from Varian treatment logs. This enables better retrospective analysis of patient motion for improved quality of care.
Area of Science:
- Medical Physics
- Radiation Oncology
- Image Analysis
Background:
- Current Varian intrafraction motion review (IMR) offers limited qualitative data at the treatment console.
- Quantitative analysis of intrafraction motion is crucial for treatment accuracy and quality improvement.
Purpose of the Study:
- To develop a method for extracting and interpreting quantitative data from combined log files for intrafraction motion review.
- To enable retrospective analysis of intrafraction motion using log file data.
Main Methods:
- Developed a parsing code to scan Varian combined log files for specific data.
- Extracted key parameters including gantry angle, marker positions, and detection results.
- Compared intrafraction motion review (IMR) data with Calypso for dual-surrogate monitoring.
Main Results:
- Successfully extracted clinically relevant quantitative parameters from combined log files.
- Established a method for dual-surrogate intrafraction monitoring by comparing IMR and Calypso data.
- Demonstrated the feasibility of quantitative intrafraction motion analysis from treatment logs.
Conclusions:
- Quantitative analysis of intrafraction motion is achievable using combined log files.
- This method enhances the review of intrafraction motion, aiding quality improvement in patient care.