Technical note: Unexpected external markers artifact in 3D k-space based parallel imaging turbo spin-echo magnetic
Frederik Crop1, Raphaëlle Mouttet-Audouard2, Xavier Mirabel2
1Medical Physics, Centre Oscar Lambret, Lille, France.
Summary
Radiotherapy MRI markers can create artifacts mimicking metastases, especially with CAIPIRINHA parallel imaging and 3D distortion correction. These imaging artifacts, distinct from compressed sensing, can lead to misdiagnosis in patients.
Area of Science:
- Medical Imaging
- Radiotherapy Physics
- Radiology
Background:
- Accurate spatial referencing is crucial for MRI-guided radiotherapy planning.
- Immobilization devices and external markers are commonly used for spatial referencing.
- Image artifacts can compromise the accuracy of radiotherapy planning.
Purpose of the Study:
- To investigate a challenging artifact in MRI for radiotherapy planning.
- To identify the cause of artifacts resembling metastases when combining specific MRI techniques and markers.
- To evaluate the impact of marker proximity and imaging parameters on artifact generation.
Main Methods:
- Utilized a 3D variable flip angle Turbo Spin Echo sequence on 1.5T and 3T MRI scanners.
- Tested two types of markers (Liquimark LM1, Spee-D-Mark) with flexible and head/neck coils.
- Employed a silicone oil phantom and 3D Fourier transforms to analyze artifact origins, comparing CAIPIRINHA and compressed sensing acceleration.
Main Results:
- Markers can appear in unexpected locations, not aligned with original or reconstructed slices, particularly with CAIPIRINHA acceleration and 3D distortion correction.
- Artifact occurrence probability increases with marker-patient distance (13% at <=2mm, 80% at >2mm).
- Artifacts were observed post-contrast, mimicking metastases, and did not occur with compressed sensing acceleration.
Conclusions:
- External markers used in radiotherapy MRI can generate artifacts misinterpreted as metastases.
- The combination of CAIPIRINHA parallel imaging and 3D distortion correction increases artifact risk.
- High signal intensity surface structures, like ears, can also cause similar diagnostic errors.