Improved evaluation of backscatter characteristics of soft tissue using high-frequency annular array
Takeru Mizoguchi1, Kenji Yoshida2, Jonathan Mamou3
1Graduate School of Science and Engineering, Chiba University, Yayoicho, Inage, Chiba 263-8522, Japan.
Abstract:
Clinical ultrasound is widely used for quantitative diagnosis. To clarify the relationship between anatomical and acoustic properties, high resolution imaging using high-frequency ultrasound (HFU) is required. However, when tissue properties are evaluated using HFU, the depth of field (DOF) is limited. To overcome this problem, an annular array transducer, which has a simple structure and produces high-quality images, is applied to HFU measurement. In previous phantom experiments, we demonstrated that the HFU annular array extends the DOF compared to that of a single-element transducer for quantitative ultrasound (QUS) analysis. Here, we extend that work by applying QUS methods to an ex vivo rat liver. The present study demonstrates that an annular array extends the region and improves the resolution for tissue characterization for an excised healthy rat liver. Amplitude envelope statistics and spectral-based analysis are used as QUS methods.