Characterisation of in-room leakage and scattered radiation for the Varian Halcyon linear accelerator
Kirstie Caravani1, Rebecca Murry2, Brendan Healy3
1Icon Cancer Centre, Toowoomba, QLD, Australia. kscara@zoho.com.
Abstract:
Unintended sources of secondary radiation resulting from photon beams in linear accelerators are patient scatter, collimator scatter, scatter from surfaces within the bunker, and head leakage. This work characterises the in-room leakage and scattered radiation for the Varian Halcyon linear accelerator. Scattered and leakage radiation for static gantry angles 0°, 45°, and 90° and largest field size 28 × 28 cm2 were measured with an ionisation chamber survey meter at a radial distance of 1.5 m from the isocentre. The scatter within the treatment room was characterised with isocontour maps from measurements of 360° arc deliveries with the largest field size 28 × 28 cm2 at various heights and distances from the isocentre. The transmission through the primary beam stopper was measured with a Farmer ionisation chamber. For static gantry angles, instantaneous dose rate readings were typically around 70 mSv/hr at 1.5 m from the isocentre with lower dose rates at room angles adjacent to the gantry. The head leakage was measured as less than 0.03% of the useful beam. In a full 360° arc, the radiation dose around the Halcyon was coldest in lateral areas, with hotter areas behind and in front of the gantry. The primary beam stopper transmission was measured as 0.019%, reducing the requirement of primary barriers in the shielding design by a factor of 1/500. The results presented in this study can be used to determine the out-of-field dose to patients and to inform bunker shielding designs for Halcyon linear accelerators.
More Related Videos
Related Concept Videos
Absorption of Radiation
Radiation: Applications
The average...
Radiation Pressure: Problem Solving
The average value of the rate of momentum transfer divided by the absorbing area represents the average force...
Conduction, Convection and Radiation: Problem Solving
In order to solve a problem related to heat transfer, first of all, the situation needs to be examined to determine the type of heat transfer involved. This could...
Lossy Lines and Overvoltages
Attenuation
When constant series resistance and shunt conductance are present, voltage and current equations are modified. The propagation constant indicates that voltage and current waves consist of both forward and backward traveling components. These waves attenuate as they propagate, with the attenuation factor related to the resistance and conductance. In a...
Boundary Conditions: Lossless Lines
At the receiving end, the boundary condition states that the voltage equals the product of the receiving-end impedance and current. This relationship is expressed as a function of the incident and...