Riata lead failure manifests as electrical storm with lead parameters showing unique postural variations: a case
Abhimanyu Uppal1, Sanjeev Kathuria1, Bhushan Shah1
1Department of Cardiology, GB Pant Institute of Post Graduate Education and Research, MAMC, JLN Road, New Delhi 110002, India.
Riata implantable cardioverter-defibrillator (ICD) leads with conductor externalization (CE) require close monitoring, as delayed electrical lead failure (ELF) can occur years later. Postural changes in lead parameters can help detect this dysfunction.
Area of Science:
- Cardiology
- Biomedical Engineering
- Medical Device Technology
Background:
- Riata implantable cardioverter-defibrillator (ICD) leads are associated with a specific mechanical failure mode: conductor externalization (CE).
- This CE can lead to delayed-onset electrical lead failure (ELF), posing a risk to patient safety.
- Understanding the long-term implications of CE in Riata ICD leads is crucial for clinical management.
Observation:
- A case study involving a 60-year-old male with ischemic cardiomyopathy and atrial fibrillation who received a Riata ICD lead.
- During a pulse generator replacement, CE was noted in the ventricular lead, but electrical parameters were initially normal.
- Four years later, the patient experienced multiple inappropriate ICD shocks due to ventricular electrogram noise, linked to postural changes in lead parameters.
Findings:
- The patient's Riata ICD lead exhibited delayed electrical lead failure (ELF) manifesting as ventricular electrogram noise and inappropriate shocks.
- Lead impedance and R-wave amplitude varied significantly with patient posture, indicating an underlying electrical issue.
- Successful resolution was achieved by implanting a new ventricular lead and abandoning the affected Riata lead.
Implications:
- Electrically inert conductor externalization (CE) in Riata ICD leads necessitates vigilant follow-up due to the potential for late electrical lead failure (ELF).
- Analyzing electrograms and monitoring lead parameters, especially with postural changes, is vital for detecting and characterizing electrical dysfunction.
- This case highlights the importance of long-term surveillance for patients with Riata ICD leads experiencing CE.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
09:16Use of a Wireless Video-EEG System to Monitor Epileptiform Discharges Following Lateral Fluid-Percussion Induced Traumatic Brain Injury
Published on: June 21, 2019
Related Concept Videos
Dysrhythmias V: Evaluating Dysrhythmias
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Dysrhythmias II: Classification of Tachyarrhythmias
Dysrhythmias III: Characteristics of Dysrhythmias