SIR-Spheres Activity Measurements Reveal Systematic Miscalibration
Stephen A Graves1,2,3, Molly Martin4, Ashok Tiwari4,5
1Department of Radiology, University of Iowa, Iowa City, Iowa; stephen-a-graves@uiowa.edu.
Independent measurements show that Yttrium-90 (90Y) SIR-Spheres contain significantly more radioactive activity than stated by the manufacturer. This suggests the calibration of these brachytherapy sources may be underestimated.
Area of Science:
- Nuclear Medicine
- Radiopharmacology
- Medical Physics
Background:
- Selective Internal Radiation Therapy (SIRT) utilizes Yttrium-90 (90Y) SIR-Spheres for liver cancer treatment.
- Accurate quantification of radioactive activity in these brachytherapy sources is crucial for effective and safe patient dosing.
- Previous calibrations by manufacturers may require independent verification.
Purpose of the Study:
- To conduct an independent and traceable activity measurement of 90Y SIR-Spheres.
- To verify the accuracy of manufacturer-provided activity calibrations for 90Y SIR-Spheres.
Main Methods:
- Utilized gamma-spectroscopic measurements with a high-purity germanium detector to analyze the 90Y internal pair production decay mode.
- Corrected raw detection rates for radioactive decay, dead time, attenuation, and geometric effects.
- Determined detection efficiency using two independent methods traceable to the National Institute of Standards and Technology (NIST).
Main Results:
- Measured activity in five SIR-Spheres vials exceeded the manufacturer's specified calibration values, with an average measured-to-calibrated activity ratio of 1.233 ± 0.030.
- Activity measurements obtained using two distinct efficiency calibration methods demonstrated excellent agreement, with results within 1% of each other.
- The findings indicate a consistent overage of radioactive activity compared to manufacturer calibrations.
Conclusions:
- The primary activity calibration for SIR-Spheres appears to significantly underestimate the true radioactive content.
- Independent verification of brachytherapy source activity is essential for ensuring accurate therapeutic doses.
- Further investigation into the calibration procedures for 90Y SIR-Spheres may be warranted.
More Related Videos
Related Concept Videos
Random and Systematic Errors
Uncertainty in Measurement: Accuracy and Precision
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...