Remote-controlled cholangiography injection device: first clinical study in China
Huan Chen1,2, Li-Yu Shan1,3, Tao Ma1
1National Local Joint Engineering Research Center for Precision Surgery and Regenerative Medicine, First Affiliated Hospital of Xi'an Jiaotong University, Xi'an, 710061, China.
Background:
X-ray cholangiography is of great value in the imaging of biliary tract diseases; however, occupational radiation exposure is unavoidable. Moreover, clinicians must manually inject the contrast dye, which may result in a relatively high incidence of adverse reactions due to unstable injection pressure. Thus, there is a need to develop a novel remote-controlled cholangiography injection device.
Methods:
Patients with external biliary drainage requiring cholangiography were included. A remote-controlled injection device was developed with three major components: an injection pump, a pressure sensor, and a wireless remote-control panel. Image quality, adverse reactions, and radiation dose were evaluated.
Results:
Different kinds of X-ray cholangiography were successfully and smoothly performed using this remote-controlled injection device in all patients. The incidence of adverse reactions in the device group was significantly lower than that in the manual group (4.17% vs. 13.9%, P = 0.001), and increasing the injection pressure increased the incidence of adverse reactions. In addition, the device helped operators avoid ionizing radiation completely.
Conclusions:
With good control of injection pressure (within 10 kPa), the remote-controlled cholangiography injection device could replace the need for the doctor to inject contrast agent with good security and effectivity. It is expected to be submitted for clinical application.